| نام تجاری: | Sinuo |
| شماره مدل: | SN881-960L |
| مقدار تولیدی: | 1 |
| شرایط پرداخت: | T/T |
| توانایی عرضه: | 10 مجموعه در هر ماه |
Climatic Test Chamber Environmental Test Chamber Laboratory Test Chamber 960L
Complied Standards & Clauses
It conforms to standards:
IEC60068-2-1:2007 “Environmental Testing-Part 2-1: Tests-Test A: Cold, IDT”
IEC60068-2-2:2007 “Environmental Testing-Part 2-1: Tests- Test B-Dry heat, IDT”
IEC60068-2-78:2012 “Environmental Testing-Part 2-78: Tests-Test Cab: Damp heat, steady state, IDT”
IEC60068-2-30:2005 “Environmental Testing-Part 2-30: Tests-Test Db: Damp heat, cyclic (12h+12h cycle), IDT”
Product Overview
The 960L Temperature Humidity Chamber is an advanced environmental testing solution designed to simulate a wide range of climatic conditions — from extreme heat and cold to varying levels of humidity and dryness. It provides a stable and precise testing environment for evaluating the performance and durability of materials and products such as electronics, electrical appliances, plastics, and other industrial components.
By exposing samples to alternating temperature and humidity cycles, this chamber helps manufacturers identify potential weaknesses, improve product design, and ensure long-term reliability during storage, transportation, and real-world use. It’s an essential tool for quality assurance and reliability testing across multiple industries.
Technical Parameters
| Technical Parameters | ||
| Control & Operation | PLC and touch screen intelligent control system | |
|
Chamber Size
|
Inner Size (mm) | External Size (mm) |
| 2000*600*800 | 2250*1650*1820 | |
| Temperature Range | -20°C ~+150°C | |
| Humidity Range |
20%~98%RH |
|
| Temperature fluctuation | ±0.5℃ | |
| Temperature deviation | ≤±2.0℃ | |
| Humidity fluctuation | ≤2.5% | |
| Humidity Deviation | ≤±3% | |
| Cooling Rate | ≧1°C/min, average no load | |
| Heating Rate | ≧2-3°C/min | |
| Working noise |
A sound level≤70dB(A) (Measured in a soundproof room with an ambient temperature of 25°C and low echo. Using A weighting, test the average of 9 points. Each test point is 1 meter away from the noise source horizontally and 1 meter above the ground.) |
|
| Load condition | None | |
| Conditions of use | |
| a) Installation site |
The ground is flat, well ventilated, and does not contain flammable, explosive, corrosive gases and dust There is no strong electromagnetic radiation source nearby. There is a drainage floor drain near the equipment (within 2 meters of the refrigeration unit). The bearing capacity of the site ground: not less than 800kg/m2. There is appropriate maintenance space around the equipment.
A: not less than 600mm B: not less than 1000mm C: not less than 100mm D: not less than 600mm |
| b) Water source requirements |
1. Purified water must be used 2. Tap water must be treated |
| c) Environmental conditions |
1. Temperature: 5℃~35℃ 2. Relative humidity: ≤85% 3. Air pressure: 86kPa~106kPa |
|
d) Power supply
|
AC380V three-phase four-wire + protective ground wire Allowable voltage fluctuation range: AC (380±10)V Allowable frequency fluctuation range: (50±0.5)Hz Protective ground wire grounding resistance is less than 4Ω The user is required to configure an air or power switch of corresponding capacity for the equipment at the installation site, and this switch must be used independently for this equipment |
| Power capacity | About 12kW |
| Maximum current | 32A |
| Power switch | 63A (molded case leakage protector) |
| e) Environmental requirements for storage equipment | When the equipment is not working, the ambient temperature should be kept within 0℃~+45℃ |
Details
![]()
![]()